A Preclinical Study of the Safety and Efficacy of Occlusin (TM) 500 Artificial Embolization Device in Sheep

作者:Owen Richard J*; Nation Patrick N; Polakowski Robert; Biliske Jennifer A; Tiege Paul B; Griffith Irwin J
来源:CardioVascular and Interventional Radiology, 2012, 35(3): 636-644.
DOI:10.1007/s00270-011-0218-7

摘要

This study evaluated the safety, effectiveness, and biodegradation of a new embolic agent, Occlusin (TM) 503 Artificial Embolization Device (OCL 503). The agent consists of biodegradable poly-lactic-co-glycolic acid microspheres (150-212 mu m) coated with type I bovine collagen and was compared with Embosphere(A (R)) Microspheres (300-500 mu m) in this controlled study of uterine artery embolization (UAE) in sheep.
Unilateral UAE was performed in 32 adult ewes randomly assigned. Vessels were embolized to effective stasis. The cohort was divided into four groups, which were sacrificed at 1, 3, 6, and 12 months.
Both agents were 100% effective in achieving stasis. At 6 months, all OCL 503-treated arteries were occluded, the microspheres degraded with time, and at 12 months all four animals examined demonstrated recanalization. OCL 503 was found in the untreated uterine artery in one animal with no other evidence of non target embolization. In the Embosphere-treated group, all vessels remained occluded and microspheres were detected in the contralateral uterine artery in 6 of 15 examined vessels and in 10 vaginal, 2 ovarian, and 1 vesical artery. No procedural-related complications were seen in either group.
OCL 503 is as effective an embolic agent as Embosphere(A (R)) Microspheres when embolizing ovine uterine arteries and resorbs with time, allowing recanalization of the treated arteries. No device-related issues or adverse events were observed.

  • 出版日期2012-6