A single-walled carbon nanotube thin film-based pH-sensing microfluidic chip

作者:Li Cheng Ai; Han Kwi Nam; Pham Xuan Hung; Seong Gi Hun*
来源:Analyst, 2014, 139(8): 2011-2015.
DOI:10.1039/c3an02195e

摘要

A novel microfluidic pH-sensing chip was developed based on pH-sensitive single-walled carbon nanotubes (SWCNTs). In this study, the SWCNT thin film acted both as an electrode and a pH-sensitive membrane. The potentiometric pH response was observed by electronic structure changes in the semiconducting SWCNTs in response to the pH level. In a microfluidic chip consisting of a SWCNT pH-sensing working electrode and an Ag/AgCl reference electrode, the calibration plot exhibited promising pH-sensing performance with an ideal Nernstian response of 59.71 mV pH(-1) between pH 3 and 11 (standard deviation of the sensitivity is 1.5 mV pH(-1), R-2 = 0.985). Moreover, the SWCNT electrode in the microfluidic device showed no significant variation at any pH value in the range of the flow rate between 0.1 and 15 mu l min(-1). The selectivity coefficients of the SWCNT electrode revealed good selectivity against common interfering ions.

  • 出版日期2014