Absolute quantification of cerebral blood flow: correlation between dynamic susceptibility contrast MRI and model-free arterial spin labeling

作者:Knutsson Linda*; van Westen Danielle; Petersen E**en T; Bloch Karin Markenroth; Holtas Stig; Stahlberg Freddy; Wirestam Ronnie
来源:Magnetic Resonance Imaging, 2010, 28(1): 1-7.
DOI:10.1016/j.mri.2009.06.006

摘要

Purpose: To compare absolute cerebral blood flow (CBF) estimates obtained by model-free arterial spin labeling (ASL) and dynamic susceptibility contrast MRI (DSC-MRI), corrected for partial volume effects (PVEs). Methods: CBF was measured using DSC-MRI and model-free ASL (quantitative signal targeting with alternating radiofrequency labeling of arterial regions) at 3 T in 15 subjects with brain tumor, and the two modalities were compared with regard to CBF estimates in normal gray matter (GM) and DSC-to-ASL CBF ratios in selected tumor regions. The DSC-MRI CBF maps were calculated using a global arterial input AIF) from the sylvian-fissure region, but, in order to minimize PVEs, the AIF time integral was rescaled by a venous output function time integral obtained from the sagittal sinus. Results: In GM, the average DSC-MRI CBF estimate was 150+/-45 ml/(min 100 g) (mean+/-SD) while the corresponding ASL CBF was 44+/-10 ml/(min 100 g). The linear correlation between GM CBF estimates obtained by DSC-MRI and ASL was r=.89, and observed DSC-to-ASL CBF ratios differed by less than 3% between GM and tumor regions. Conclusions: A satisfactory positive linear correlation between the CBF estimates obtained by model-free ASL and DSC-MRI was observed, and DSC-to-ASL CBF ratios showed no obvious tissue dependence.

  • 出版日期2010-1