Direct quantitative analysis of HCV RNA by atomic force microscopy without labeling or amplification

作者:Jung Yu Jin; Albrecht Jeffrey A; Kwak Ju Won; Park Joon Won*
来源:Nucleic Acids Research, 2012, 40(22): 11728-11736.
DOI:10.1093/nar/gks953

摘要

Force-based atomic force microscopy (AFM) was used to detect HCV (hepatitis C virus) RNA directly and to quantitatively analyse it without the need for reverse transcription or amplification. Capture and detection DNA probes were designed. The former was spotted onto a substrate with a conventional microarrayer, and the latter was immobilized on an AFM probe. To control the spacing between the immobilized DNAs on the surface, dendron self-assembly was employed. Force-distance curves showed that the mean force of the specific unbinding events was 32 +/- 5 pN, and the hydrodynamic distance of the captured RNA was 30-60 nm. Adhesion force maps were generated with criteria including the mean force value, probability of obtaining the specific curves and hydrodynamic distance. The maps for the samples whose concentrations ranged from 0.76 fM to 6.0 fM showed that cluster number has a linear relationship with RNA concentration, while the difference between the observed number and the calculated one increased at low concentrations. Because the detection limit is expected to be enhanced by a factor of 10 000 when a spot of 1 micron diameter is employed, it is believed that HCV RNA of a few copy numbers can be detected by the use of AFM.

  • 出版日期2012-12