MRI Mode Programming for Safe Magnetic Resonance Imaging in Patients With a Magnetic Resonance Conditional Cardiac Device

作者:Nakai Toshiko*; Kurokawa Sayaka; Ikeya Yukitoshi; Iso Kazuki; Takahashi Keiko; Sasaki Naoko; Ashino Sonoko; Okubo Kimie; Okumura Yasuo; Kunimoto Satoshi; Watanabe Ichiro; Hirayama Atsushi
来源:International Heart Journal, 2016, 57(2): 173-176.
DOI:10.1536/ihj.15-284

摘要

Although diagnostically indispensable, magnetic resonance imaging (MRI) has been, until recently, contraindicated in patients with an implantable cardiac device. MR conditional cardiac devices are now widely used, but the mode programming needed for safe MRI has yet to be established. We reviewed the details of 41 MRI examinations of patients with a MR conditional device. There were no associated adverse events. However, in 3 cases, paced beats competed with the patient's own beats during the MRI examination. We describe 2 of the 3 specific cases because they illustrate these potentially risky situations: a case in which the intrinsic heart rate increased and another in which atrial fibrillation occurred. Safe MRI in patients with an MR conditional device necessitates detailed MRI mode programming. The MRI pacing mode should be carefully and individually selected.

  • 出版日期2016-3