An MRI Receiver Coil Produced by Inkjet Printing Directly on to a Flexible Substrate

作者:Mager Dario; Peter Andreas; Del Tin Laura; Fischer Elmar; Smith Patrick J*; Hennig Juergen; Korvink Jan G
来源:IEEE Transactions on Medical Imaging, 2010, 29(2): 482-487.
DOI:10.1109/TMI.2009.2036996

摘要

Inkjet printing has been used to produce resonant radio frequency coils that are comparable to those produced by conventional printed circuit board (PCB) methods. The coils, which consist of a conductive loop and in-series capacitors, form part of a receiver circuit that is used for magnetic resonance imaging (MRI). The resonant circuit is selective at the predetermined frequency of 400 MHz. The required electrical components (resistor, capacitor, and inductor) were produced by inkjet printing, with scaling experiments for resistor and capacitor performed before the complete loops with integrated capacitors were printed. Numerical simulation was used to determine the required values for the components. The inkjet printed circuit was combined with a small tuning and matching board before being connected to a network analyzer and the MRI hardware. With a matching of 38 dB at 400 MHz the achieved results were comparable to those from standard PCB techniques. The performance of the inkjet printed component as a receiver device for nuclear magnetic resonance and MRI was verified by imaging reference phantoms and a whole kiwifruit; it compares favorably to standard MRI devices. Inkjet printing can, therefore, be considered a feasible technique for producing MRI receiver circuits on flexible substrates.

  • 出版日期2010-2