Analytical Performance of the Roche LightCycler (R) Mycobacterium Detection Kit for the Diagnosis of Clinically Important Mycobacterial Species

作者:Omar Shaheed V*; Roth Andreas; Ismail Nazir A; Erasmus Linda; Ehlers Marthie; Kock Marleen; Paulse Nuraan; Said Halima M; Hoosen Anwar A; Reischl Udo
来源:PLos One, 2011, 6(9): e24789.
DOI:10.1371/journal.pone.0024789

摘要

Background: The LightCycler (R) Mycobacterium Detection Kit based on real-time PCR technology for the detection of Mycobacterium tuberculosis, Mycobacterium avium and Mycobacterium kansasii was recently developed. This study evaluated its analytical sensitivity, specificity and reproducibility.
Methodology/Principal Findings: Plasmid standards were prepared and used to determine the limit of detection. The assay was also performed against organisms other than mycobacteria, other mycobacterial strains and interfering substances to exclude cross-reactivity and interference. Reference standards were prepared and tested to assess the assay's reproducibility. All PCR assays were performed using the LightCycler (R) 2.0 Instrument. The detection limit for M. tuberculosis was 28 copies per microlitre. Neither cross-reactivity nor interference occurred with non-mycobacterial organisms and substances tested. Overall reproducibility for consecutive measurements, run-to-run, lot-to-lot, day-to-day and laboratory-to-laboratory achieved a coefficient of variance of less than two percent.
Significance: The LightCycler (R) Mycobacterium Detection kit has shown to be a robust and accurate assay with the potential to be used as a rapid TB diagnostic test.