摘要

N-nitrosamines are probable human carcinogens that could seriously affect the safety of drinking water consumers. In this study, a rapid, sensitive and reliable solid-phase extraction method combined with ultra-performance liquid chromatography coupled with tandem mass spectrometry has been developed for the quantification of N-nitrosodiphenylamine and N-nitrosodimethylamine at ultratrace levels in aqueous matrices. Chromatographic separation was performed using an Acquity UPLC BEH C18 column and a mobile phase consisting of acetonitrile, water, and formic acid (60:40:0.1, v/v/v) at a flow rate of 0.4 mL min(-1). The run time of the instrumental method was two minutes. Under optimized conditions the performance of the proposed method was studied in terms linearity (r(2) >= 0.998), precision (< 3%), accuracy (between 99% and 103%), lower limit of detection (0.04-0.16 ng L-1), and lower limit of quantification (0.08-0.32 ng L-1). The obtained extraction recoveries of the target compounds were within the range of 85-101%, and the relative standard deviations were less than 4%. The matrix effect was within 84-99% at all quality control levels. These results showed clearly that, the developed method is rapid, sensitive efficient, and accurate for simultaneous determination of N-nitrosodiphenylamine and N-nitrosodimethylamine. In addition, the proposed method has been successfully applied for the screening of these molecules in real water samples.

  • 出版日期2015-2