Minimum maximum temperature gradient coil design

作者:While Peter T*; Poole Michael S; Forbes Larry K; Crozier Stuart
来源:Magnetic Resonance in Medicine, 2013, 70(2): 584-594.
DOI:10.1002/mrm.24492

摘要

Ohmic heating is a serious problem in gradient coil operation. A method is presented for redesigning cylindrical gradient coils to operate at minimum peak temperature, while maintaining field homogeneity and coil performance. To generate these minimaxT coil windings, an existing analytic method for simulating the spatial temperature distribution of single layer gradient coils is combined with a minimax optimization routine based on sequential quadratic programming. Simulations are provided for symmetric and asymmetric gradient coils that show considerable improvements in reducing maximum temperature over existing methods. The winding patterns of the minimaxT coils were found to be heavily dependent on the assumed thermal material properties and generally display an interesting fish-eye spreading of windings in the dense regions of the coil. Small prototype coils were constructed and tested for experimental validation and these demonstrate that with a reasonable estimate of material properties, thermal performance can be improved considerably with negligible change to the field error or standard figures of merit. Magn Reson Med 70:584-594, 2013.

  • 出版日期2013-8