A New Device for Intraoperative Renal Blood Flow Measurement During Open-Heart Surgery: An Experimental Study and the Clinical Pilot Study

作者:Tirilomis Theodor; Popov Aron F; Hanekop Gunnar G; Braeuer Anselm; Quintel Michael; Schoendube Friedrich A; Friedrich Martin G*
来源:Artificial Organs, 2013, 37(10): 927-932.
DOI:10.1111/aor.12084

摘要

Renal blood flow (RBF) may vary during cardiopulmonary bypass and low flow may cause insufficient blood supply of the kidney triggering renal failure postoperatively. Still, a valid intraoperative method of continuous RBF measurement is not available. A new catheter combining thermodilution and intravascular Doppler was developed, first calibrated in an in vitro model, and the catheter specific constant was determined. Then, application of the device was evaluated in a pilot study in an adult cardiovascular population. The data of the clinical pilot study revealed high correlation between the flow velocities detected by intravascular Doppler and the RBF measured by thermodilution (Pearson%26apos;s correlation range: 0.78 to 0.97). In conclusion, the RBF can be measured excellently in real time using the new catheter, even under cardiopulmonary bypass.

  • 出版日期2013-10