A noncontact intraocular pressure measurement device using a micro reflected air pressure sensor for the prediagnosis of glaucoma

作者:Kim Kyoung Hwan*; Kim Byeong Hee; Seo Young Ho
来源:Journal of Micromechanics and Microengineering, 2012, 22(3): 035022.
DOI:10.1088/0960-1317/22/3/035022

摘要

This study investigates a novel, portable tonometer using a micro reflected air pressure sensor for the prediagnosis of glaucoma. Because glaucoma progresses slowly and is not painful, glaucoma patients require a portable prediagnosis system to periodically measure intraocular pressure at home. Conventionally, intraocular pressure is measured by an air-puff tonometer whereby the cornea is deformed by a short pulse of air pressure and the magnitude of the corneal deformation is measured by optic systems such as a combination of laser- and photodiodes. In this study, a micro reflected air pressure sensor was designed, fabricated, and tested in order to measure the magnitude of corneal deformation without optic systems. In an experimental study, artificial eyes with different internal pressures were fabricated and these pressures were measured by the aforementioned system.

  • 出版日期2012-3