Accuracy of single-plane fluoroscopy in determining relative position and orientation of total knee replacement components

作者:Acker Stacey*; Li Rebecca; Murray Heather; John Paul St; Banks Scott; Mu Shang; Wyss Urs; Deluzio Kevin
来源:Journal of Biomechanics, 2011, 44(4): 784-787.
DOI:10.1016/j.jbiomech.2010.10.033

摘要

The accuracy of estimating the relative pose between knee replacement components, in terms of clinical motion, is important in the study of knee joint kinematics. The objective of this study was to determine the accuracy of the single-plane fluoroscopy method in calculating the relative pose between the femoral component and the tibial component, along knee motion axes, while the components were in motion relative to one another. The kinematics of total knee replacement components were determined in vitro using two simultaneous methods: single-plane fluoroscopic shape matching and an optoelectronic motion tracking system. The largest mean differences in relative pose between the two methods for any testing condition were 2.1 degrees, 0.3 degrees, and 1.1 degrees in extension, abduction, and internal rotation respectively, and 1.3, 0.9, and 1.9 mm in anterior, distal, and lateral translations, respectively. For the optimized position of the components during dynamic trials, the limits of agreement, between which 95% of differences can be expected to fall, were -2.9 to 4.5 degrees in flexion, -0.9 to 1.5 degrees in abduction, -2.4 to 2.1 degrees in external rotation, -2.0 to 3.9 mm in anterior-posterior translation, -2.2 to 0.4 mm in distal-proximal translation and -7.2 to 8.6 mm in medial-lateral translation. These mean accuracy values and limits of agreement can be used to determine whether the shape-matching approach using single-plane fluoroscopic images is sufficiently accurate for an intended motion tracking application.

  • 出版日期2011-2-24