An investigation of medical radiation detection using CMOS image sensors in smartphones

作者:Kang Han Gyu; Song Jae Jun; Lee Kwonhee; Nam Ki Chang; Hong Seong Jong; Kim Ho Chul*
来源:Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment , 2016, 823: 126-134.
DOI:10.1016/j.nima.2016.04.007

摘要

Medical radiation exposure to patients has increased with the development of diagnostic X-ray devices and multi-channel computed tomography (CT). Despite the fact that the low-dose CT technique can significantly reduce medical radiation exposure to patients, the increasing number of CT examinations has increased the total medical radiation exposure to patients. Therefore, medical radiation exposure to patients should be monitored to prevent cancers caused by diagnostic radiation. However, without using thermoluminescence or glass dosimeters, it is hardly measure doses received by patients during medical examinations accurately. Hence, it is necessary to develop radiation monitoring devices and algorithms that are reasonably priced and have superior radiation detection efficiencies. The aim of this study is to investigate the feasibility of medical dose measurement using complementary metal oxide semiconductor (CMOS) sensors in smartphone cameras with an algorithm to extract the X-ray interacted pixels. We characterized the responses of the CMOS sensors in a smartphone with respect to the X-rays generated by a general diagnostic X-ray system. The characteristics of the CMOS sensors in a smartphone camera, such as dose response linearity, dose rate dependence, energy dependence, angular dependence, and minimum detectable activity were evaluated. The high energy gamma-ray of 662 keV from Cs-137 can be detected using the smartphone camera. The smartphone cameras which employ the developed algorithm can detect medical radiations.

  • 出版日期2016-7-1