Scanning electrochemical microscopy at variable temperatures

作者:Schaefer Dominik; Puschhof Andrea; Schuhmann Wolfgang*
来源:Physical Chemistry Chemical Physics, 2013, 15(14): 5215-5223.
DOI:10.1039/c3cp43520b

摘要

All chemical reactions are influenced by temperature, however, temperature is usually not considered an important parameter which has to be varied or at least controlled in SECM measurements. A precise temperature-control unit was designed and integrated into a SECM setup which allows setting the temperature of the sample and the adjacent electrolyte in a range between 0 and 100 degrees C without causing convection. Data acquisition was synchronized with the current pulses through the Peltier element to decrease the noise and keep the tip-to-sample distance constant during imaging. SECM images in the feedback mode, generator collector mode and the redox competition mode for model samples such as an enzyme entrapped within a polymer spot or oxygen reduction catalysts demonstrate the importance of controlling temperature as well as performing SECM experiments at predefined and constant increased temperature.

  • 出版日期2013