Measured response of a liquid scintillation detector to quasi-monoenergetic electrons and neutrons

作者:Rout P C; Gandhi A; Basak T; Thomas R G; Ghosh C; Mitra A; Mishra G; Behera S P; Kujur R; Mirgule E T; Nayak B K; Saxena A; Kumar S; Datar V M
来源:Journal of Instrumentation, 2018, 13(01): P01027.
DOI:10.1088/1748-0221/13/01/P01027

摘要

The response of a liquid scintillator (EJ-301) to monoenergetic electrons tagged by Compton back scattered gamma-rays has been measured using various radioactive gamma-ray sources. It is observed that the measured electron response is linear up to an energy of similar to 4 MeV. The resolution of the liquid scintillator at an energy 1 MeVee is found to be similar to 11%. The pulse shape discrimination and pulse height response of the liquid scintillator for neutrons have been measured using the Li-7(p,n(1))Be-7*(0.429 MeV) reaction. A nonlinear response to mono-energetic neutrons for the liquid scintillator is observed at energies E-n = 5.3 MeV, 9.0 MeV and 12.7 MeV. In addition, the measured response of the liquid scintillator for electrons and neutrons has been compared with simulated results obtained using the Monte Carlo based Geant4 toolkit.

  • 出版日期2018-1