A carbon monoxide and thermometry sensor based on mid-IR quantum-cascade laser wavelength-modulation absorption spectroscopy

作者:Vanderover J; Wang W; Oehlschlaeger M A*
来源:Applied Physics B: Lasers and Optics , 2011, 103(4): 959-966.
DOI:10.1007/s00340-011-4570-8

摘要

A mid-infrared wavelength modulation spectroscopy (WMS) absorption sensor utilizing a quantum cascade laser (QCL) source for carbon monoxide detection at 4.6 mu m and high-temperature thermometry has been developed and demonstrated. Measurements are made in a room-temperature gas cell and in high-temperature reflected-shock-heated gases. Room-temperature results demonstrate a minimum CO detectivity of 0.03 ppm per meter of absorption path length at a 1-kHz detection bandwidth, representing an order-of-magnitude increased sensitivity compared to scanned-wavelength direct absorption. High-temperature shock tube measurements demonstrate sensitive thermometry from 850 to 3500 K for pressures near 1 atm at detection bandwidths of 500 Hz and 1 kHz. Sensor determined temperatures agree with the post-reflected-shock conditions within +/- 1.9% (1 sigma deviation). To our knowledge, the thermometry measurements presented here represent the first WMS high-temperature thermometry measurements made using a QCL.

  • 出版日期2011-6