In this paper we proposed a reverse high performance liquid chromatography method for the simultaneous determination of three N-methylcarbamates (NMCs) named carbofuran, carbaryl and methiocarb, using the post-column chemiluminescence (CL) detection with the luminol reaction. This method is based on the enhancing effect of these analytes on the CL emission generated by the oxidation of luminol with potassium permanganate in alkaline medium. The separation was reached in less than 14 min using a C18 column and an isocratic binary mobile phase consisting of acetonitrile:water (50:50, v/v) pumped at a flow rate of 1 mL min(-1). CL reagents (luminol and KMnO(4)) were incorporated by means of a peristaltic pump and were firstly mixed using a three-way connector. Then this stream was mixed with the eluate using another three-way connector just before reaching the detection cell. The optimization of variables affecting the CL reaction (reaction medium, concentration, flow rate of reagents and distance between both connectors) were optimized by means of experimental designs. Ethiofencarb, a NMC which has nowadays fallen into disuse was used as internal standard. For the analysis of theses pesticides in real water samples a pre-treatment step consisting of solid phase extraction (SPE) was conducted in order to reach sensitivity levels below the legal maximum concentration permitted. In the case of vegetable sample, SPE was used for matrix cleaning purpose.