摘要

A new polyaniline carbon paste electrode prepared by mixing polyaniline (emeraldine), nafion, graphite powder and urease for urea analysis was exploited. The ratio of polyaniline, nafion, urease and graphite for the construction of the electrodes and the optimal conditions for urea determination were studied. The detection limit of this sensor for urea is 5 mu M and the linearity from 5 mu M to 7.5 mM is obtained in FIA. This sensor has a response time of 90s and shows good reproducibility and stability (RSD, 6.3%, n = 43). The blood samples from a patient during blood dialysis were taken and analyzed. The urea concentrations in blood obtained from this sensor are comparable with urea test kit method.