A high blood level in the venous chamber and a wet-stored dialyzer help to reduce exposure for microemboli during hemodialysis

作者:For**erg Ulf*; Jonsson Per; Stegmayr Christofer; Jonsson Fredrik; Nilsson Bo; Nilsson Ekdahl Kristina; Stegmayr Bernd
来源:Hemodialysis International, 2013, 17(4): 612-617.
DOI:10.1111/hdi.12052

摘要

During hemodialysis (HD), microemboli develop in the blood circuit of the apparatus. These microemboli can pass through the venous chamber and enter into the patient%26apos;s circulation. The aim of this study was to investigate whether it is possible to reduce the risk for exposure of microemboli by altering of the treatment mode. Twenty patients on chronic HD were randomized to a prospective cross-over study of three modes of HD: (a) a dry-stored dialyzer (F8HPS, Fresenius, steam sterilized) with a low blood level in the venous chamber (DL), (b) the same dialyzer as above, but with a high level in the venous chamber (DH), and (c) a wet-stored dialyzer (Rexeed, Asahi Kasei Medical, gamma sterilized) with a high blood level (WH). Microemboli measurements were obtained in a continuous fashion during 180 minutes of HD for all settings. A greater number of microemboli were detected during dialysis with the setting DL vs. WH (odds ratio [OR] 4.07, 95% confidence interval [CI] 4.03-4.11, P%26lt;0.0001) and DH vs. WH (OR 1.18, 95% CI 1.17-1.19, P%26lt;0.0001) and less for DH vs. DL (OR 0.290, 95% CI 0.288-0.293, P%26lt;0.0001). These data indicate that emboli exposure was least when using WH, greater with DH, and most with DL. This study shows that using a high blood level in the venous chamber and wet-stored dialyzers may reduce the number of microemboli.

  • 出版日期2013-10