A novel polytetrafluoroethylene-channel model, which simulates low levels of culturable bacteria in buildup biofilm after repeated endoscope reprocessing

作者:Alfa Michelle J; Ribeiro Maira M; Luciano Cristiana da Costa; Franca Rodrigo; Olson Nancy; DeGagne Pat; Singh Harminder
来源:Gastrointestinal Endoscopy, 2017, 86(3): 442-U350.
DOI:10.1016/j.gie.2017.05.014

摘要

Background and Aims: Clinical studies have shown variable culture results from flexible endoscope channels possibly because of low levels of bacteria that are difficult to extract. The aim of this study was to develop a simulated-use buildup biofilm (BBF) model that mimics low levels of viable bacteria after repeated rounds of aldehyde fixation and accumulation. Methods: New endoscope channels were exposed to 8 days of repeated rounds of biofilm formation using ATS2015 containing Enterococcus faecalis and Pseudomonas aeruginosa, rinsing, fixation with glutaraldehyde, and rinsing. Viable count and scanning electron microscopy and borescope examination were used to compare the impact of dry storage over 26 weeks on the level of culturable bacteria and to compare the Centers for Disease Control and Prevention flush method of channel harvesting with a flush-brush-flush method. Results: E faecalis (log(10) 6.6) and P aeruginosa (log(10) 8.6) accumulated over 8 days of cyclic biofilm formation and partial glutaraldehyde fixation, but after a final exposure to 2.6% glutaraldehyde the level of culturable bacteria was less than 2 log(10). The Centers for Disease Control and Prevention channel harvesting method appeared by borescope to be inferior to a flush-brush-flush sample collection method for detection of viable bacteria. P aeruginosa increased up to 7 log(10) after 26 weeks of dry storage, indicating there were viable but nonculturable bacteria present initially that recovered during storage. Conclusions: Viable but nonculturable P aeruginosa within the BBF model are able to recover, and this phenomenon may explain the variability of culture in patient-used endoscopes. Our data also indicated that friction may be a critical part of sample collection from endoscope channels.

  • 出版日期2017-9