A quick wireless label-free detection of disease-related C-reactive proteins (CRPs) using a 200-mu m-long micro-electromechanical systems (MEMS) microcantilever housed in a 7 x 7 mm(2) reaction chamber with a safe reusable feature is reported. The assay time ranges from about 30 min to 3 h, depending on accuracy. The deflection of the microcantilever due to specific CRP-antiCRP binding is detected using a position-sensitive detector. The converted bio-signal is transmitted by a custom designed wireless amplitude-shift-keying (ASK) transceiver IC fabricated in a 0.18 mu m CMOS process. CRP concentrations from 1 mu g/mL to 500 mu g/mL can be detected. A 0.2-Hz I-V ac signal instead of traditional bases/acids is applied to the bio-MEMS sensor to unbind the CRP from the microcantilever for reusability.