A COMPARISON STUDY OF RADIATION EXPOSURE TO PATIENTS DURING EVAR AND DYNACT IN AN ANGIOSUITE VS. AN OPERATING THEATRE

作者:Bruschi Andrea*; Michelagnoli Stefano; Chisci Emiliano; Mazzocchi Silvia; Panci Simone; Didona Annamaria; Ghirelli Alessandro; Zatelli Giovanna
来源:Radiation Protection Dosimetry, 2015, 163(4): 491-498.
DOI:10.1093/rpd/ncu228

摘要

The aim of this study was to assess the patient dosimetric impact of endovascular abdominal aortic aneurysm repair (EVAR), both in an operating theatre (OR) and in an angiosuite (AS), with the facility of Dynamic CT (Dyna CT, Siemens AG, Berlin, Germany). One hundred and forty-six consecutive EVAR procedures dating from May 2011 to March 2013 were analysed. These were performed either in an OR (n = 97) using a mobile C-arm or in an AS (n = 49) equipped with a ceiling-mounted angiography system. Air kerma area product (P-KA) and total air kerma at reference point (K-a,K-r) values were reported for all procedures and Dyna CT. Radiation exposure during EVAR was quite low in the majority of patients but nearly 50 % higher if performed in AS vs. OR. Median Dyna CT K-a,K-r was the same as an entire EVAR procedure in OR. The higher patient's radiation exposure recorded in the AS should be balanced with the technical advantages given to the EVAR procedure.

  • 出版日期2015-3