Accuracy of Human and Veterinary Point-of Care Glucometers for Use in Rhesus Macaques (Macaca mulatta), Sooty Mangabeys (Cercocebus atys), and Chimpanzees (Pan troglodytes)

作者:Clemmons Elizabeth A; Stovall Melissa I; Owens Devon C; Scott Jessica A; Jones Wilkes Amelia C; Kempf Doty J; Ethun Kelly F*
来源:Journal of the American Association for Laboratory Animal Science, 2016, 55(3): 346-353.

摘要

Handheld, point-of-care glucometers are commonly used in NHP for clinical and research purposes, but whether these devices are appropriate for use in NHP is unknown. Other animal studies indicate that glucometers should be species-specific, given differences in glucose distribution between RBC and plasma; in addition, Hct and sampling site (venous compared with capillary) influence glucometer readings. Therefore, we compared the accuracy of 2 human and 2 veterinary glucometers at various Hct ranges in rhesus macaques (Macaca mulatta), sooty mangabeys (Cercocebus atys), and chimpanzees (Pan troglodytes) with that of standard laboratory glucose analysis. Subsequent analyses assessed the effect of hypoglycemia, hyperglycemia, and sampling site on glucometer accuracy. The veterinary glucometers overestimated blood glucose (BG) values in all species by 26 to 75 mg/dL. The mean difference between the human glucometers and the laboratory analyzer was 7 mg/dL or less in all species. The human glucometers overestimated BG in hypoglycemic mangabeys by 4 mg/dL and underestimated BG in hyperglycemic mangabeys by 11 mg/dL; similar patterns occurred in rhesus macaques. Hct did not affect glucometer accuracy, but all samples were within the range at which glucometers generally are accurate in humans. BG values were significantly lower in venous than capillary samples. The current findings show that veterinary glucometers intended for companion-animal species are inappropriate for use in the studied NHP species, whereas the human glucometers showed clinically acceptable accuracy in all 3 species. Finally, potential differences between venous and capillary BG values should be considered when comparing and evaluating results.

  • 出版日期2016-5