摘要

A simple signal conditioning circuit using a transformer ratio arm (TRA) bridge for converting capacitance change into frequency for capacitive sensors is presented. The circuit employs a relaxation oscillator in which the output frequency is linearly related to the capacitive unbalance of a TRA bridge. The design, analysis, and experimental results of the circuit and its application to a thin-film-based humidity sensor are reported. The experimental results confirm the theoretical value predicted. The circuit which offers the minimum parasitic earth capacitance effect has the potential for accurately monitoring measurement parameters, particularly ppm-level humidity. The simulation results for the effect of parasitic earth capacitances and ambient temperature on the output frequency have also been discussed. The pulse wave output of the circuit is interfaced with microcontroller for direct moisture display in ppm. The frequency sensitivity and nonlinearity of the sensor for the 0-110-ppm moisture range are found to be 10.94 Hz/ppm and 1.2%, respectively.

  • 出版日期2015-12