An array of virtual Frisch-grid CdZnTe detectors and a front-end application-specific integrated circuit for large-area position-sensitive gamma-ray cameras

作者:Bolotnikov A E*; Ackley K; Camarda G S; Cherches C; Cui Y; De Geronimo G; Fried J; Hodges D; Hossain A; Lee W; Mahler G; Maritato M; Petryk M; Roy U; Salwen C; Vernon E; Yang G; James R B
来源:Review of Scientific Instruments, 2015, 86(7): 073114.
DOI:10.1063/1.4927455

摘要

We developed a robust and low-cost array of virtual Frisch-grid CdZnTe detectors coupled to a front-end readout application-specific integrated circuit (ASIC) for spectroscopy and imaging of gamma rays. The array operates as a self-reliant detector module. It is comprised of 36 close-packed 6 x 6 x 15 mm(3) detectors grouped into 3 x 3 sub-arrays of 2 x 2 detectors with the common cathodes. The front-end analog ASIC accommodates up to 36 anode and 9 cathode inputs. Several detector modules can be integrated into a single-or multi-layer unit operating as a Compton or a coded-aperture camera. We present the results from testing two fully assembled modules and readout electronics. The further enhancement of the arrays' performance and reduction of their cost are possible by using position-sensitive virtual Frisch-grid detectors, which allow for accurate corrections of the response of material non-uniformities caused by crystal defects.

  • 出版日期2015-7