An Integrated UV-Vis Microfluidic Device for the Study of Concentrated Solvent Extraction Reactions

作者:Ciceri Davide*; Nishi Kazuiko; Sievens Geoffrey W; Perera Jilska M
来源:Solvent Extraction Research and Development-Japan, 2013, 20(0): 197-203.
DOI:10.15261/serdj.20.197

摘要

A custom built microfluidic device that potentially enables real time monitoring of solute transfer across a water/oil interface is presented and discussed. The device consisted of a Pyrex (R) glass, Y-Y shaped microchannel fabricated with an inbuilt UV-Vis detection system. The device was connected to a spectrograph by means of optical fibres for detection of the signal. The device was tested against both a blue coloured dye (waxoline blue) and the metallorganic complex of Co(II) and di (2-ethylhexyl) phosphoric acid (DEHPA). Due to the very short optical path length (similar to 10(-4) cm), in situ measurements of highly concentrated analytes are shown to be possible. The use of the device for measurements of concentrated solvent extraction kinetics is envisaged.

  • 出版日期2013