Development of double-sector positron emission tomography detector for tumour tissue detection

作者:Qi Hui Rong*; Xiao Guo Qing; Wang Jin Chuan; Guo Zhong Yan; Zhan Wen Long; Xu Zhi Guo; Wu Li Jie; Ding Xian Li; Gao Hui; Xu Hu Shan; Li Jia Xing; Sun Zhi Yu; Hu Zheng Guo; Mao Rui Shi; Li Chen; Chen Li Xin; Zhao Tie Cheng
来源:Atomic Energy Science and Technology, 2005, 39(4): 345-348.

摘要

With the establishment of heavy ion cooling storage ring (CSR) in Institute of Modern Physics, Chinese Academy of Sciences, the DPET (double-sector positron emission tomography) detector is developed. The structure of DPET detector, the principle of detection and characteristic of design are particularly discussed, including the structure of detection crystals, the optical fibers, the design of the signal readout circuit and the electronics data acquisition system. Some results of detector unit performance test are also presented.

  • 出版日期2005-7

全文