Assessing agreement in measurements of orthodontic study models: Digital caliper on plaster models vs 3-dimensional software on models scanned by structured-light scanner

作者:Hassan Wan Nurazreena Wan*; Othman Siti Adibah; Chan Chee Seng; Ahmad Roshahida; Ali Siti Nor'Ain; Rohim Anis Abd
来源:American Journal of Orthodontics and Dentofacial Orthopedics, 2016, 150(5): 886-895.
DOI:10.1016/j.ajodo.2016.04.021

摘要

Introduction: In this study we aimed to compare measurements on plaster models using a digital caliper, and on 3-dimensional (3D) digital models, produced using a structured-light scanner, using 3D software. Methods: Fifty digital models were scanned from the same plaster models. Arch and tooth size measurements were made by 2 operators, twice. Calibration was done on 10 sets of models and checked using the Pearson correlation coefficient. Data were analyzed by error variances, repeatability coefficient, repeated-measures analysis of variance, and Bland-Altman plots. Results: Error variances ranged between 0.001 and 0.044 mm for the digital caliper method, and between 0.002 and 0.054 mm for the 3D software method. Repeated-measures analysis of variance showed small but statistically significant differences (P < 0.05) between the repeated measurements in the arch and buccolingual planes (0.011 and 0.008 mm, respectively). There were no statistically significant differences between methods and between operators. Bland-Altman plots showed that the mean biases were close to zero, and the 95% limits of agreement were within +/- 0.50 mm. Repeatability coefficients for all measurements were similar. Conclusions: Measurements made on models scanned by the 3D structured-light scanner were in good agreement with those made on conventional plaster models and were, therefore, clinically acceptable.

  • 出版日期2016-11