Mini-implant stability at the initial healing period A clinical pilot study

作者:Nienkemper Manuel*; Wilmes Benedict; Pauls Alexander; Drescher Dieter
来源:The Angle Orthodontist, 2014, 84(1): 127-133.
DOI:10.2319/040813-271.1

摘要

Objective: To evaluate the changes of mini-implant stability over the initial healing period in humans. Material and Methods: A sample of 19 consecutively treated patients (mean age 15.5 +/- 7.3 years) was examined. In each patient, a mini-implant of a size of 2 x 9 mm was inserted into the anterior palate. Implant stability was assessed using resonance frequency analysis (RFA) immediately after insertion (T0), 2 weeks later (T1), 4 weeks later (T2), and 6 weeks later (T3). Insertion depth (ID) and the maximum insertion torque (IT) were measured. Data were tested for correlations between RFA, ID, and IT. All RFA values were tested for statistically significant differences between the different times. Results: The mean ID was 7.5 +/- 0.6 mm, and the mean IT was 16.8 +/- 0.6 Ncm. A correlation was found between RFA and ID (r = .726, P < .0001), whereas no correlations between RFA and IT or between IT and ID were observed. From T0 to T1, the stability (36.1 +/- 6.1 implant stability quotient [ISQ]) decreased nonsignificantly by 4.9 +/- 6.1 ISO values (P > .05). Between T1 and T2, the stability decreased highly significantly (P < .001) by 7.9 +/- 5.9 ISQ values. From T2 on, RFA remained nearly unchanged (-1.7 +/- 3.5 ISO; P > .05). Conclusions: Mini-implant stability is subject to changes during the healing process. During weeks 3 and 4, a significant decrease of the stability was observed. After 4 weeks, the stability did not change significantly.

  • 出版日期2014-1