摘要

The authors describe a fluorometric assay for the determination of Hg(II). A naphthalimide derivative is used as a label for a thymine (T) rich ssDNA, and graphene oxide magnetized with Fe3O4 nanoparticles acts as a quencher and preconcentrators. In the absence of Hg(II), the labeled ssDNA does not separate from the magnetized graphene oxide. As a result, fluorescence is fully quenched. In the presence of Hg(II), a T-Hg(II)-T link is formed dues to the highly affinity between T and Hg(II). Hence, fluorescence is restored. The assay has a linear response in the 1.0 to 10.0 nM Hg(II) concentration range, and a 0.65 nM detection limit. The method is selective and sensitive. It was applied to the analysis of spiked environmental water samples, and data agreed well with those obtained by atomic fluorescence spectrometry.