摘要

Despite environmental and geochemical interests, Cr and Fe have been left beyond the reach of determinations by ICP-MS due to severe interferences originating from Ar. The applicability of a dynamic reaction cell (DRC)-ICP-MS has been examined for determinations in environmental and geochemical samples. The reaction with NH(3) in the DRC system provides an eligible technique to determine Cr, because of a greater improvement in the signal/noise (SIN) ratio due to an effective elimination of interferences arising from Ar (ArC, ArN and ArO), and makes it possible to analyze Cr even at sub-mu g L(-1) levels. As compared to non-DRC mode analyses, the DRC technique using m/z 56 appeared to be preferable for Fe determination in most terrestrial waters because of effective suppression of (40)Ar(16)O(+). In addition, the effects of cluster ions, such as (39)K(14)N(1)H(3)(+) and (40)Ca(14)N(1)H(2)(+), on Fe determination were also negligibly small. Measurements using (54)Fe by the DRC mode are also advantageous for Ca-rich samples, such as limestone and dolomite.

  • 出版日期2010-8