Use of Normothermic Default Humidifier Settings Causes Excessive Humidification of Respiratory Gases During Therapeutic Hypothermia

作者:Tanaka Shoichiro; Iwata Sachiko; Kinoshita Masahiro; Tsuda Kennosuke; Sakai Sayaka; Saikusa Mamoru; Shindo Ryota; Harada Eimei; Okada Junichiro; Hisano Tadashi; Kanda Hiroshi; Maeno Yasuki; Araki Yuko; Ushijima Kazuo; Sakamoto Teruo; Yamashita Yushiro; Iwata Osuke*
来源:Therapeutic Hypothermia and Temperature Management, 2016, 6(4): 180-188.
DOI:10.1089/ther.2016.0005

摘要

Adult patients frequently suffer from serious respiratory complications during therapeutic hypothermia. During therapeutic hypothermia, respiratory gases are humidified close to saturated vapor at 37 degrees C (44mg/L) despite that saturated vapor reduces considerably depending on temperature reduction. Condensation may cause serious adverse events, such as bronchial edema, mucosal dysfunction, and ventilator-associated pneumonia during cooling. To determine clinical variables associated with inadequate humidification of respiratory gases during cooling, humidity of inspiratory gases was measured in 42 cumulative newborn infants who underwent therapeutic hypothermia. Three humidifier settings of 37-default (chamber outlet, 37 degrees C; distal circuit, 40 degrees C), 33.5-theoretical (chamber outlet, 33.5 degrees C; distal circuit, 36.5 degrees C), and 33.5-adjusted (optimized setting to achieve 36.6mg/L using feedback from a hygrometer) were tested to identify independent variables of excessively high humidity >40.7mg/L and low humidity <32.9mg/L. The mean (SD) humidity at the Y-piece was 39.2 (5.2), 33.3 (4.1), and 36.7 (1.2) mg/L for 37-default, 33.5-theoretical, and 33.5-adjusted, respectively. The incidence of excessive high humidity was 10.3% (37-default, 31.0%; 33.5-theoretical, 0.0%; 33.5-adjusted, 0.0%), which was positively associated with the use of a counter-flow humidifier (p<0.001), 37-default (compared with 33.5-theoretical and 33.5-adjusted, both p<0.001) and higher fraction of inspired oxygen (p=0.003). The incidence of excessively low humidity was 17.5% (37-default, 7.1%; 33.5-theoretical, 45.2%; 33.5-adjusted, 0.0%), which was positively associated with the use of a pass-over humidifier and 33.5-theoretical (both p<0.001). All patients who used a counter-flow humidifier achieved the target gas humidity at the Y-piece (36.6 +/- 0.5mg/L) required for 33.5-adjusted with 33.5-theoretical. During cooling, 37-default is associated with excessively high humidity, whereas 33.5-theoretical leads to excessively low humidity. Future studies are needed to assess whether a new regimen with optimized Y-piece temperature and humidity control reduces serious respiratory adverse events during cooling.

  • 出版日期2016-12