3D-CONFORMAL VERSUS INTENSITY-MODULATED POSTOPERATIVE RADIOTHERAPY OF VAGINAL VAULT: A DOSIMETRIC COMPARISON

作者:Cilla Savino; Macchia Gabriella*; Digesu Cinzia; Deodato Francesco; Romanella Michele; Ferrandina Gabriella; Padula Gilbert D A; Picardi Vincenzo; Scambia Giovanni; Piermattei Angelo; Morganti Alessio Giuseppe
来源:Medical Dosimetry, 2010, 35(2): 135-142.
DOI:10.1016/j.meddos.2009.04.003

摘要

We evaluated a step-and-shoot IMRT plan in the postoperative irradiation of the vaginal vault compared with equispaced beam arrangements (3-5) 3D-radiotherapy (RT) optimized plans. Twelve patients were included in this analysis. Four plans for each patient were compared in terms of dose-volume histograms, homogeneity index (HI), and conformity index (CI): (1) 3 equispaced beam arrangement 3D-RT; (2) 4 equispaced beam arrangement 3D-RT; (3) 5 equispaced beam arrangement 3D-RT; (4) step-and-shoot IMRT technique. CI showed a good discrimination between the four plans. The mean scores of CI were 0.58 (range: 0.38-0.67) for the 3F-CRT plan, 0.58 (range: 0.41-0.66) for 4F-CRT, 0.62 (range: 0.43-0.68) for 5F-CRT and 0.69 (range: 0.58-0.78) for the IMRT plan. A significant improvement of the conformity was reached by the IMRT plan (p < 0.001 for all comparisons). As expected, the increment of 3D-CRT fields was associated with an improvement of target dose conformity and homogeneity; on the contrary, in the IMRT plans, a better conformity was associated to a worse target dose homogeneity. A significant reduction in terms of D(mean), V90%, V95%, V100% was recorded for rectal and bladder irradiation with the IMRT plan. Surprisingly, IMRT supplied a significant dose reduction also for rectum and bladder V30% and V50%. A significant dosimetric advantage of IMRT over 3D-RT in the adjuvant treatment of vaginal vault alone in terms of treatment conformity and rectum and bladder sparing is shown.

  • 出版日期2010