摘要

To date, no emerging preclinical or clinical near-infrared fluorescence (NIRF) imaging devices for noninvasive and/or surgical guidance have their performances validated on working standards with SI units of radiance that enable comparison or quantitative quality assurance. In this work, we developed and deployed a methodology to calibrate a stable, solid phantom for emission radiance with International System of Units (SI) units of for use in characterizing the measurement sensitivity of ICCD and IsCMOS detection, signal-to-noise ratio, and contrast. In addition, at calibrated radiances, we assess transverse and lateral resolution of ICCD and IsCMOS camera systems. The methodology allowed demonstration of superior SNR of the ICCD over the IsCMOS technology and superior resolution of the IsCMOS over the ICCD approach. Contrast depended upon the camera settings (binning and integration time) and gain of intensifier. Finally, because the architecture of CMOS and CCD camera systems results in vastly different performance, we comment on the utility of these technologies for small animal imaging as well as clinical applications for noninvasive and surgical guidance.

  • 出版日期2016-3