Ambient temperature impacts on pH of exhaled breath condensate

作者:Koczulla Andreas R*; Noeske Sarah; Herr Christian; Dette Frank; Pinkenburg Olaf; Schmid Severin; Joerres Rudolf A; Vogelmeier Claus; Bals Robert
来源:Respirology, 2010, 15(1): 155-159.
DOI:10.1111/j.1440-1843.2009.01664.x

摘要

Background and objective: Analysis of exhaled breath condensate (EBC) pH is a non-invasive method to study airway inflammation. Low pH is correlated with inflammatory diseases like asthma and COPD. The aim of this study was to assess the influence of measurement temperature on pH values of EBC. Methods: EBC was collected using the RTube in 10 healthy non-smoking controls, 10 smokers before and after cigarette smoking, 10 stable COPD patients and 10 patients with exacerbated COPD. pH was determined directly after degassing at temperatures of 23 degrees C and 37 degrees C. Results: When comparing all groups pH was significantly (P = 0.0002) higher (mean +/- SD 7.88 +/- 0.92) at 37 degrees C as compared with 23 degrees C (7.44 +/- 0.90). Specifically, at 23 degrees C pH was significantly lower in the group of exacerbated COPD (6.78 +/- 1.27) and healthy non-smoking controls (8.04 +/- 0.39). In contrast, subgroup analysis of values assessed at 37 degrees C did not display significant differences. Conclusions: Our data indicate a considerable influence of temperature on pH values in EBC. Thus the temperature at which pH measurements in EBC studies are performed should be declared.

  • 出版日期2010-1