摘要

The technique of SPECT-RNV (radionuclide ventriculography) offers a greater amount of clinically usable data than its planar counterpart (P-RNV). In transitioning from planar to SPECT-only acquisition methodologies, reprojection of the SPECT data can provide a planar dataset which can be used as an interim technique. The aim of this study was to test if reprojected planar images could be used as a surrogate for true planar images in SPECT-only setting. We performed SPECT-RNV and P-RNV on 47 patients on traditional sodium iodide (NaI) cameras, determining left ventricular ejection fractions (LVEF) for planar (EFP) and SPECT (EFS) techniques. We reprojected the SPECT-RNV data along the best septal separation angle determined from planar scanning. This creates a further planar dataset denoted 'reprojected P-RNV' (rP-RNV) giving a reprojected ejection fraction (EFR) which can be used as a validation variable in transitioning to SPECT-only acquisition. Performing t tests showed no statistical difference between EFP and EFR (P > .017) but bias was observed in EFS results compared to EFP and EFS compared to EFR results. An unblinded, comparison of parametric data between the three datasets for a subset of ten patients showed good clinical concordance. False negative and false positive rates were low for rP-RNV compared to P-RNV. The reprojected planar LVEF correlates well to P-RNV EF values. The rP-RNV dataset can aid clinicians in transitioning from planar RNV to SPECT-only acquisition.

  • 出版日期2014-10