Microfabricated Gas Chromatograph for Rapid, Trace-Level Determinations of Gas-Phase, Explosive Marker Compounds

作者:Collin William R; Serrano Gustavo; Wright Lindsay K; Chang Hungwei; Nunovero Nicolas; Zellers Edward T*
来源:Analytical Chemistry, 2014, 86(1): 655-663.
DOI:10.1021/ac402961t

摘要

A prototype microfabricated gas chromatograph (mu GC) adapted specifically for the rapid determination of selected gas-phase marker compounds of the explosive 2,4,6-trinitrotoluene (TNT) at sub-parts-per-billion (< ppb) concentrations in complex mixtures is described. Si-microfabricated focuser, separation column, and sensor array components are integrated with a high-volume sampler of conventional construction to reduce analysis time and the limit of detection (LOD). The primary markers selected as target analytes were 2,4-dinitrotoluene (2,4-DNT, a persistent impurity of TNT) and 2,3-dimethyl-2,3-dinitrobutane (DMNB, a taggant), along with 2,6-dinitrotoluene (2,6-DNT, a lessprominent TNT impurity), which was also included in numerous tests. Selective preconcentration, on-column focusing, temperature-programmed chromatographic separation, and sensor array detection/recognition facilitated determinations of the primary markers in the presence of 20 (or more) interferences within similar to 2 min under laptop control. Estimated LODs of 2.2, 0.48, and 0.86 ng were achieved for DMNB, 2,6-DNT, and 2,4-DNT, respectively, which correspond to 0.30, 0.067, and 0.12 ppb in each 1-L air sample collected.

  • 出版日期2014-1-7