摘要

A method is proposed for plotting calibration graphs by reference analyte solutions (tandem calibration) for the analysis of steels and alloys by inductively coupled plasma atomic emission spectrometry (ICP-AES) with spark ablation. The use of spectral lines of analytes and an internal standard with similar values of sums of ionization and excitation energy ensures an excellent repeatability and accuracy of the results of analysis using tandem calibration. A nebulizer chamber for the simultaneous introduction of solutions and solid sample aerosols into inductively coupled plasma (ICP) is designed. The optimal parameters of the introduction of sample aerosols and a method of plasma observation are chosen; the requirements for the choice of the analytical lines of analytes are proposed.

  • 出版日期2013-7

全文