摘要

Objective: To compare 2 open-ear hearing solutions for sloping high-frequency sensorineural hearing loss: open-fit hearing aid (HA) and active middle ear implant (AMEI).
Study Design: Within-subjects prospective design.
Setting: Tertiary referral hospital.
Patients and Devices: Fourteen patients with sloping, high-frequency sensorineural hearing loss were recruited from 39 patients previously implanted with an AMEI and 10 agreed to participate, ranging in age from 44 to 73 years ( mean, 59 yr). Patients were selected because their hearing thresholds (500-3,000 Hz) qualified them for both AMEI and open-fit HA use. All patients received a Vibrant Soundbridge ( Vibrant MED-EL) for an average of 25.1 months before data collection and used their AMEI on a daily basis. The open-fit HA used in this study was the Delta 8000 (Oticon). Both study devices have been fit with the specific fitting strategies as recommended by the manufacturer.
Main Outcome Measures: Sound-field hearing thresholds, Freiburger monosyllabic words in quiet, speech reception thresholds for 50% correct recognition for Freiburger numbers and for Oldenburg sentences in quiet, and speech reception thresholds for 50% correct recognition for Oldenburg sentences in noise.
Results: Both HA and AMEI conditions showed significantly better sound-field thresholds and speech recognition on monosyllabic word and sentence tests in quiet and in noise than in the unaided condition. A comparison of aided conditions revealed that, in the AMEI-aided condition, high-frequency audibility and speech discrimination scores in quiet and in noise were significantly better than those in the open-fit HA.
Conclusion: Both open-fit HAs and AMEIs provided audiologic benefit to patients with sloping high-frequency sensorineural hearing loss. However, despite overlapping indication criteria for the 2 devices, performance with the AMEI was significantly better for the AMEI than for the open-fit HA.