AMBIENT DOSE ESTIMATION H*(10) FROM LaBr3(Ce) SPECTRA

作者:Camp A; Vargas A*
来源:Radiation Protection Dosimetry, 2014, 160(4): 264-268.
DOI:10.1093/rpd/nct342

摘要

The stripping method for ambient dose estimation has been used for detectors such as high-purity Ge (HPGe). This method strips the spectrum from the partial absorptions produced in the detector leaving only the events corresponding to the full absorption of a gamma ray. In the present study, this method is applied to a 1 %26apos;%26apos; x 1 %26apos;%26apos; LaBr3(Ce) detector using the PENELOPE/penEasy Monte Carlo code to obtain both the partial absorptions and detector full peak efficiency. The stripping method has been validated from a set of gamma fluxes carried out at the accredited laboratory of the Institute of Energy Technologies of the Technical University of Catalonia and results were obtained with differences %26lt;5 %. After validation, the LaBr3(Ce) monitor was installed on the roof of the institute premises working in parallel with a photon equivalent dose monitor, model FHZ 601A from the FAG Company. The derived H*(10) values from the LaBr3(Ce) detector show good agreement with those derived from the dose monitor.

  • 出版日期2014-8