A NATIONAL PROGRAMME FOR PATIENT AND STAFF DOSE MONITORING IN INTERVENTIONAL CARDIOLOGY

作者:Sanchez R*; Vano E; Fernandez J M; Sotil J; Carrera F; Armas J; Rosales F; Pifarre X; Escaned J; Angel J; Diaz J F; Bosa F; Saez J R; Goicolea J
来源:Radiation Protection Dosimetry, 2011, 147(1-2): 57-61.
DOI:10.1093/rpd/ncr275

摘要

A national programme on patient and staff dose evaluation in interventional cardiology made in cooperation with the haemodynamic section of the Spanish Society of Cardiology has recently been launched. Its aim is to propose a set of national diagnostic reference levels (DRLs) for patients as recommended by the International Commission on Radiological Protection and to initiate several optimisation actions to improve radiological protection of both patients and staff. Six hospitals have joined the programme and accepted to submit their data to a central database. First to be acquired were the quality control data of the X-ray systems and radiation doses of patients and professionals. The results from 9 X-ray systems, 1467 procedures and staff doses from 43 professionals were gathered. Provisional DRLs resulted in 44 Gy cm(2) for coronary angiography and 78 Gy cm2 for interventions. The X-ray systems varied up to a factor of 5 for dose rates in reference conditions. Staff doses showed that 50 % of interventional cardiologists do not use their personal dosemeters correctly.

  • 出版日期2011-9