Development of high-resolution 4D in vivo-CT for visualization of cardiac and respiratory deformations of small animals

作者:Sera Toshihiro*; Yokota Hideo; Fujisaki Kazuhiro; Fukasaku Kazuaki; Tachibana Hiroyuki; Uesugi Kentaro; Yagi Naoto; Himeno Ryutaro
来源:Physics in Medicine and Biology, 2008, 53(16): 4285-4301.
DOI:10.1088/0031-9155/53/16/005

摘要

The interest in small animal models of human diseases has generated a need to design a computed tomography (CT) system that operates at a microscopic level. It is particularly important to be able to visualize the dramatic rhythmical motion of organs such as the heart and lungs. In order to evaluate the motion of the heart and lungs of small animals (rats and mice), we developed in the present study a high-resolution 4D in vivo-CT system for small animals that uses synchrotron radiation. To reduce motion artifacts and the radiation dose, the projections were synchronized with airway pressure, the ECG, the x-ray shutter and the CCD shutter. For cardiovascular imaging, a blood pool contrast agent was injected and the data sets were acquired at several ECG points during the end- expiratory phase. For imaging of the lungs, the data sets were acquired at several airway pressures during diastole. The dynamic motion of the cardiovascular system (the ventricles and coronary arteries) and small airways (diameter > 250 mu m of rats and 125 mu m of mice) was visualized. This high-resolution imaging tool may be very useful for the development of novel drugs in murine models, in addition to its use in the study of cardiovascular and respiratory physiology.

  • 出版日期2008-8-21