Automated laboratory based X-ray beamline with multi-capillary sample chamber

作者:Purushothaman S; Gauthe B L L E; Brooks N J; Templer R H; Ces O*
来源:Review of Scientific Instruments, 2013, 84(8): 085104.
DOI:10.1063/1.4816825

摘要

An automated laboratory based X-ray beamline with a multi-capillary sample chamber capable of undertaking small angle X-ray scattering measurements on a maximum of 104 samples at a time as a function of temperature between 5 and 85 degrees C has been developed. The modular format of the system enables the user to simultaneously equilibrate samples at eight different temperatures with an accuracy of +/- 0.005 degrees C. This system couples a rotating anode generator and 2D optoelectronic detector with Franks X-ray optics, leading to typical exposure times of less than 5 min for lyotropic liquid crystalline samples. Beamline control including sample exchange and data acquisition has been fully automated via a custom designed LabVIEW framework.