摘要

The signal current from an ionisation chamber with a PMMA build-up cap decreases with irradiation time due to electric fields produced by positive charges induced on the cap. In the present study, it was confirmed that the signal current decreases faster for irradiation using narrower Co-60 gamma-ray beams. This is because the number of secondary electrons that are emitted from surrounding materials and penetrate the build-up cap is smaller in a narrower gamma-ray beam, so that fewer positive charges are neutralised. The ionisation chamber was first subjected to continuous gamma-ray irradiation for 24 h, following which it was irradiated with shorter periodic gamma-ray bursts while measuring the current signal. This allowed the coefficients of positive charge accumulation and dissipation to be determined. It was found that the dissipation coefficient has a large constant value during gamma-ray irradiation and decreases asymptotically to a small value after irradiation is stopped. From the coefficients, the minimum signal current was calculated, which is the value when accumulation and dissipation balance each other under continuous irradiation. The time required for the signal current to recover following irradiation was also calculated.