Monitoring In Vivo Absorption of a Drug-Eluting Bioabsorbable Stent With Intravascular Ultrasound-Derived Parameters A Feasibility Study

作者:Bruining Nico*; de Winter Sebastiaan; Roelandt Jos R T C; Regar Evelyn; Heller Iddo; van Domburg Ron T; Hamers Ronald; Onuma Yoshinobu; Dudek Darius; Webster Mark W I; Thuesen Leif; Ormiston John A; Cheong Wai Fung; Miguel Hebert Karine; Veldhof Susan; Serruys Patrick W
来源:JACC: Cardiovascular Interventions , 2010, 3(4): 449-456.
DOI:10.1016/j.jcin.2010.02.004

摘要

Objectives The aim of this study was to investigate the feasibility of using quantitative differential echogenicity to monitor the in vivo absorption process of a drug-eluting poly-l-lactic-acid (PLLA) bioabsorbable stent (BVS, Abbott Vascular, Santa Clara, California).
Background A new bioabsorbable, balloon-expanded coronary stent was recently evaluated in a first-in-man study. Little is known about the absorption process in vivo in diseased human coronary arteries.
Methods In the ABSORB (Clinical Evaluation of the BVS everolimus eluting stent system) study, 30 patients underwent treatment with the BVS coronary stent system and were examined with intracoronary ultrasound (ICUS) after implantation, at 6 months and at 2-year follow-up. Quantitative ICUS was used to measure dimensional changes, and automated ICUS-based tissue composition software (differential echogenicity) was used to quantify plaque compositional changes over time in the treated regions.
Results The BVS struts appeared as bright hyperechogenic structures and showed a continuous decrease of their echogenicity over time, most likely due to the polymer degradation process. In 12 patients in whom pre-implantation ICUS was available, at 2 years the percentage-hyperechogenic tissue was close to pre-implantation values, indicating that the absorption process was either completed or the remaining material was no longer differentially echogenic from surrounding tissues.
Conclusions Quantitative differential echogenicity is a useful plaque compositional measurement tool. Furthermore, it seems to be valuable for monitoring the absorption process of bioabsorbable coronary stents made of semi-crystalline polymers.

  • 出版日期2010-4