A Simple and Facile Glucose Biosensor Based on Prussian Blue Modified Graphite String

作者:Lee Seung Ho; Chung Joo Ho; Park Hun Kuk; Lee Gi Ja*
来源:Journal of Sensors, 2016, 2016: 1859292.
DOI:10.1155/2016/1859292

摘要

This work describes the string sensor for the simple and sensitive detection of glucose which is based on Prussian blue (PB) modified graphite utilizing dipping. First, PB modified graphite (PB-G) strings are characterized by physical and electrochemical techniques to optimize the PB-G layer thickness. Then, glucose oxidase (GOx) is immobilized on PB-G string electrode with biocompatible chitosan overlayer (Chi/GOx/PB-G). The Chi/GOx/PB-G string electrode exhibits a sensitivity of 641.3 mu A.mM(-1).cm(-2) to glucose with a linear range of 0.03 to 1.0mM (R-2 = 0.9957) and a rapid response time (<3 s). Moreover, the Chi/GOx/PB-G string electrodes are less sensitive to common interference materials such as ascorbic acid, uric acid, galactose, and acetaminophen than to glucose. The Chi/GOx/PB-G string electrodes also show excellent reproducibility (<5% RSD). Therefore, our Chi/GOx/PB-G string electrodes can be simple, robust, and reliable tools for glucose sensing which can avoid complicated and difficult multistep fabrication processes. In addition, we expect that they have many potential applications in fields ranging from health care to food analysis, in particular where single use is favorable.

  • 出版日期2016