摘要

Rapid identification of bacterial pathogens from clinical specimens is essential to establish an adequate empirical antibiotic therapy to treat urinary tract infections (UTIs). We used matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) combined with UF-1000i urine flow cytometry of urine specimens to quickly and accurately identify bacteria causing UTIs. We divided each urine sample into three aliquots for conventional identification, UF-1000i, and MALDI-TOF MS, respectively. We compared the results of the conventional method with those of MALDI-TOF MS combined with UF-1000i, and discrepancies were resolved by 16S rRNA gene sequencing. We analyzed 1456 urine samples from patients with UTI symptoms, and 932 (64.0%) were negative using each of the three testing methods. The combined method used UF-1000i to eliminate negative specimens and then MALDI-TOF MS to identify the remaining positive samples. The combined method was consistent with the conventional method in 1373 of 1456 cases (94.3%), and gave the correct result in 1381 of 1456 cases (94.8%). Therefore, the combined method described here can directly provide a rapid, accurate, definitive bacterial identification for the vast majority of urine samples, though the MALDI-TOF MS software analysis capabilities should be improved, with regard to mixed bacterial infection.