Objective. This study aimed (1) to determine whether scanner arm rotation causes significant movement of the head restraint and (2) to measure patient movement and its variation during the scan. %26lt;br%26gt;Study Design. The iCAT scanner and a high-speed camera were used. The 40 patients were divided into 2 groups: the open-eyed group and the blindfolded group. %26lt;br%26gt;Results. The mean level of head restraint movement was 0.130 mm, with a significantly higher level at the beginning, probably owing to the accelerating arm. Mean movement of patients was 1.135 mm and 1.119 mm in the open-eyed and blindfolded groups, respectively. Patient movement was also significantly higher at the beginning of the scan, when noise and vibrations are likely to surprise the patient. %26lt;br%26gt;Conclusions. Patient instruction and a dry-run scan should be done by clinicians. Manufacturers should consider separating the seat and head restraint from the rest of the scanner to avoid vibration transfer.