DETERMINING WHEN ENHANCED PAUSE (Penh) IS SENSITIVE TO CHANGES IN SPECIFIC AIRWAY RESISTANCE

作者:Frazer David G*; Reynolds Jeffrey S; Jackson Mark C
来源:Journal of Toxicology and Environmental Health Part A: Current Issues , 2011, 74(5): 287-295.
DOI:10.1080/15287394.2010.514235

摘要

Penh is a dimensionless index normally used to evaluate changes in the shape of the airflow pattern entering and leaving a whole-body flow plethysmograph as an animal breathes. The index is sensitive to changes in the distribution of area under the waveform during exhalation and increases in a nonlinear fashion as the normalized area increases near the beginning of the curve. Enhanced pause (Penh) has been used to evaluate changes in pulmonary function and as a method to evaluate airway reactivity. However, the use of Penh to assess pulmonary function has been challenged (Bates et al., 2004; Lundblad et al., 2002; Mitzner et al., 2003; Mitzner & Tankersley, 1998; Petak et al., 2001; Sly et al., 2005). The objective of this study was to show how Penh of the thorax and plethysmograph flow patterns are related. That relationship is used to describe the conditions under which whole-body plethysmograph Penh measurements can be used to detect changes in sR(aw).

  • 出版日期2011

全文