A fully reconfigurable biopotential sensing amplifier utilizing floating-gate transistors is presented in this paper. By using the complementary differential pairs along with the current reuse technique, the theoretical limit for the noise efficiency factor of the proposed amplifier is below 1.5. Without consuming any extra power, floating-gate transistors are employed to program the low-frequency cutoff corner of the amplifier and to implement the common-mode feedback. A concept proving prototype chip was designed and fabricated in a 0.35 mu m CMOS process occupying 0.17 mm(2) silicon area. With a supply voltage of 2.5 V, the measured midband gain is 40.7 dB and the measured input-referred noise is 2.8 mu V-rms. The chip was tested under several configurations with the amplifier bandwidth being programmed to 100 Hz, 1 kHz, and 10 kHz. The measured noise efficiency factors in these bandwidth settings are 1.96, 2.01, and 2.25, respectively, which are among the best numbers reported to date. The measured common-mode rejection and the supply rejection are above 70 dB. When the bandwidth is configured to be 10 kHz, the dynamic range measured at 1 kHz is 60 dB with total harmonic distortion less than 0.1%. The proposed amplifier is also demonstrated by recording electromyography (EMG), electrocardiography (ECG), electrooculography (EOG), and electroencephalography (EEG) signals from human bodies.